1.2.6.7. SpectroDrive

1.2.6.7. SpectroDrive
1

Measure track length

1. Measure track length
Starts measuring the rail length and stores it afterwards to the instrument.
2

Slope Calibration

2. Slope Calibration
Instrument calibration to be used with a color reference chart or to adjust density values to a 3rd party instrument.
3

Scan parameter setup

3. Scan parameter setup
4

Sheet orientation

4. Sheet orientation
See details here: 1.2.6.10. Sheet Orientation
5

Generation

5. Generation
Generation 2 is for the latest one called SpectroDrive.NG, includes M1 reading
Generation 1 is for the predecessor with M0 only reading
6

Serial number

6. Serial number
Serial number of the currently connected SpectroDrive.
Use the unlock button "x" to disconnect InkZoneMove from the instrument.
7

WI-FI Channel

7. WI-FI Channel
Change here from the automatic channel selection to a fixed channel.
See details in the troubleshooting section: 1.2.6.7. SpectroDrive
8

Measurement condition

8. Measurement condition
See here the instrument measurement conditions: 1.2.6.11. Measurement Conditions M0 - M1 - M2 - M3
9

IZQuatro setup

9. IZQuatro setup
10

Troubleshooting

10. Troubleshooting
If you experience continuously connection problem to the device than try using a fixed WLAN channel.
 
Do like:
1. Download a WLAN channel scanner app to your mobile
 
2.  Find the channel with the lowest traffic
 
 
3. Change in the configuration "Wireless channel" from Automatic to a fixed channel. Channel 2 to 11 are valid.
4. Quit InkZoneMove and unplug and replug the Techkon USB WLAN box. Restart InkZoneMove. Finish
 
 
 
11

Specifications

11. Specifications
Producer
Techkon
 
Min. patch size
3.0mm
Size
20" to 80"
Connected by
WLAN module
Measurement Conditions
M0, M1, M2, M3
Comment
supports M1 with SpectroDriveNG
12

Return to Scanning Device Setup

12. Return to Scanning Device Setup